The Intoxilyzer 5000 has had numerous issues over the years. However, the government always falls back on the position that the machine runs a self diagnostic check to ensure that it is working properly. This is fine if the self diagnostic test actually works. Dr. Stephen Rose and Charles Smith, International Forensic Research Institute located at Florida International University ran various diagnostic tests on the Intoxilyzeer 5000 with key parts disconnected. They found that the machine does not ahave the ability to recognize when key parts are disconnected and still showed the diagnostic test to be OK. Some of these key parts were vital to ensure that outside interferents would not affect the test.

LIST OF ITEMS THAT CAN BE DISCONNECTED-INTOXILYZER 5000


INTOXILYZER 5000- Model 6600

1. External breath tube Diagnostic OK
2. RFI external antenna Diagnostic OK
3. RFI connector on the mother board Diagnostic OK
4. Cooling fan Diagnostic OK
5. Internal breath tube heater Diagnostic OK
6. Solenoid #1 Diagnostic OK
7. Solenoid # 2 Diagnostic OK
8. Sample chamber heater Diagnostic OK
9. Pressure switch Diagnostic OK
10. Internal pump Diagnostic OK
11. Key Board Diagnostic OK
12. LPT (DATA) port Diagnostic OK
13. Reset switch Diagnostic OK
14. Start button Diagnostic OK
15. Display Diagnostic OK
16. External breath tub heater Diagnostic OK
17. Dip switches Diagnostic OK
18. Partial Power supply Diagnostic OK
19. Printer pin control Diagnostic OK
20. I.R. Source Diagnostic OK
21. Filter wheel motor Diagnostic OK
22. Power supply No response
23. Printer head controller Processor error
24. Thermostat assembly Processor error
25. Unplugged from wall No response

The machine used was model 66 each part was disconnected singular, all other parts were still connected. The device gave the above listed messages after the self diagnostic program was run. All tests were conducted under the supervision of Dr. Stephen Rose and Charles Smith, International Forensic Research Institute located at Florida International University